2019-05-02 15:37:16 | 1ST REVIEW: |
| 2017 FLORIDA BUILDING CODE 6TH EDITION |
| 2017 FLORIDA BUILDING CODE - PLUMBING, SIXTH EDITION |
| 2017 FLORIDA BUILDING CODE - ACCESSIBILITY, SIXTH |
| EDITION |
| 2017 FLORIDA BUILDING CODE - FUEL GAS, SIXTH EDITION |
| 2017 FLORIDA BUILDING CODE - RESIDENTIAL, SIXTH EDITION |
| 2017 FLORIDA BUILDING CODE - EXISTING BUILDING, SIXTH |
| EDITION |
| 2017 FLORIDA BUILDING CODE - ENERGY CONSERVATION, SIXTH |
| EDITION |
| |
| MED GAS COMMENTS: |
| |
| 1. PLEASE PLACE NOTE ON PLAN THE CONTRACTOR AND ANYONE |
| INSTALLING MED GAS PIPING SHALL BE MED GAS CERTIFIED. |
| INSTALLERS SHALL PROVIDED PROOF FOR PERMIT ISSUE AND ON |
| JOBSITE FOR INSPECTORS APPROVAL . PER FAC 61G4-15.031 |
| |
| 2. PLEASE PROVIDE PIPING DETAIL FOR NEW ZONE VALVES. |
| PER WPB AMEND TO FBC 107.2.1 |
| |
| 3. PROVIDE AMOUNT OF NITROGEN AND NITROUS OXIDE BEING |
| STORED. PER WPB AMEND TO FBC 107.2.1 |
| |
| 4. PROVIDE DETAIL FOR CONNECTION OF THE TANK CHAINS. |
| PER WPB AMEND TO FBC 107.2.1 |
| |
| 5. PROVIDE MANUFACTURE SPECIFICATION AND INSTILLATION |
| GUIDE FOR THE MED GAS EQUIPMENT. PER WPB AMEND TO FBC |
| 107.2.1 |
| |
| |
| PLEASE NOTE THAT SUBMITTAL OF ADDITIONAL AND/OR REVISED |
| MATERIALS MAY RESULT IN NEW PLAN REVIEW COMMENTS. |
| WHEN RESUBMITTING, IT IS HELPFUL TO PROVIDE A RESPONSE |
| LETTER ADDRESSING EACH ITEM ALONG WITH THE CITY |
| RE-SUBMITTAL FORM. PLEASE, ADDITIONALLY, INSERT |
| CORRECTED PAGES INTO TO SUBMITTAL AND REMOVE OR VOID |
| THE PREVIOUSLY REVIEWED SHEETS. |
| |