Plan Review Notes For Permit 18091284 |
Permit Number |
18091284 |
|
Review Stop |
MEDGAS |
Sequence Number |
4 |
|
Notes |
Date |
Text |
2019-10-04 12:21:59 | PLUMBING PROVISO : 18091284 | | | | MED GAS COMMENTS: SHALL BE COMPLIED WITH PRIOR TO FINAL | | INSPECTION. | | | | 1) SHOW THE TERMINATION POINT OF THE VACUUM PUMP VENT | | LINE. THE DISCHARGE SHALL BE A MINIMUM OF 25 FEET FROM | | ANY DOOR, WINDOW, AIR INTAKE, OR OTHER OPENINGS IN | | BUILDINGS. NFPA 99 SEC.5.1.3.7.6.2. | | | | 2) THE MEDICAL AIR INTAKE SHALL BE LOCATED A MINIMUM OF | | 25 FEET FROM VENTILATING SYSTEM EXHAUSTS, PLUMBING | | VENTS, VACUUM DISCHARGES, OR OTHER AREAS THAT CAN | | COLLECT OTHER NOXIOUS FUMES. NFPA 99 SEC. 5.1.3.6.3.11. | | | | DEFERRED SUBMITTALS. SUBMIT MANUFACTURER'S | | SPECIFICATIONS FOR THE VACUUM PUMPS AND AIR | | COMPRESSORS, THE MUFFLER/INTAKE AND ALSO FOR THE | | RECEIVERS, DRYERS, AFTERCOOLERS, ETC.IF THEY ARE BEING | | REPLACED WITH NEW. WPB AMEND. TO FBC SEC. 107.3.4.1 | | | | | | GEORGE JOHNSON | | PLUMBING PLANS EXAMINER | | CITY OF WEST PALM BEACH | | 561-805-6711 | | [email protected] |
|