Plan Review Notes For Permit 18081039
Permit Number 18081039
Review Stop MEDGAS
Sequence Number 2
Notes
Date Text
2018-12-21 07:32:582ND REVIEW:
 2017 FLORIDA BUILDING CODE 6TH EDITION
 2017 FLORIDA BUILDING CODE - PLUMBING, SIXTH EDITION
 2017 FLORIDA BUILDING CODE - ACCESSIBILITY, SIXTH
 EDITION
 2017 FLORIDA BUILDING CODE - FUEL GAS, SIXTH EDITION
 2017 FLORIDA BUILDING CODE - RESIDENTIAL, SIXTH EDITION
 2017 FLORIDA BUILDING CODE - EXISTING BUILDING, SIXTH
 EDITION
 2017 FLORIDA BUILDING CODE - ENERGY CONSERVATION, SIXTH
 EDITION
  
 PREVIOUS MED GAS COMMENTS: 2ND REQUEST
 1. NOT COMPLIED: PROVIDE A SCHEDULE FOR ALL MED GAS
 EQUIPMENT. PER WPB AMEND TO FBC 107.2.1
 2. NOT COMPLIED: PROVIDE MANUFACTURE SPECIFICATIONS AND
 INSTALLATION GUIDES FOR ALL MED GAS EQUIPMENT INCLUDING
 THE TANKS AND VALVES. PER WPB AMEND TO FBC 107.2.1
 3. COMPLIED
 4. NOT COMPLIED: SHT. P-903 ON THE ISOMETRIC WHEN A
 PIPE IS CROSSING ANOTHER PIPE IT SHOULD HAVE A GAP AS
 THEY LOOK NOW A NUMBER OF PIPES APPEAR TO BE CONNECTED
 TOGETHER IT IS DIFFICULT TO DETERMINE WHICH SYSTEMS IS
 WHICH PLEASE CLARIFY. PER WPB AMEND TO FBC 107.2.1
 ISOMETRIC IS STILL TO CONGESTED TO READ PLEASE EXPAND.
 5. NOT COMPLIED: PLEASE PROVIDE DETAIL OF THE STRAPPING
 OF THE MED GAS STORAGE TANKS. PER WPB AMEND TO FBC
 107.2.1
 6. NOT COMPLIED: A SEPARATE PERMIT IS REQUIRED FOR MED
 GAS, ALONG WITH THE APPLICATION PROOF OF MED GAS
 CERTIFICATION FOR THE CONTRACTOR AND INSTALLER IS
 REQUIRED. PER FAC 61G4-15.031(1) (4) (A) (B) PLEASE
 PLACE NOTE ON PLAN.
 7. COMPLIED
 8. COMPLIED
 NEW COMMENTS:
 1. PLEASE PROVIDE DETAILS FOR THE MED GAS ZONE VALVES 5
 PORTS AND 2 PORTS. PER WPB AMEND TO FBC 107.2.1
 2. THE VACUUM EXHAUST PIPE ON SHEET P-802 DETAIL #4
 SHOWS THE PIPING AT 2" THE NOTE STATES IT 1-1/2" SHEET
 P-205 KEY NOTE #2 STATES IT AT STATES IT AT 3" PLEASE
 CHECK ALL NOTES DETAILS AND ISOMETRIC AND SIZE THE
 SYSTEM THE SAME PER THE MANUFACTURE SPECIFICATIONS. PER
 WPB AMEND TO FBC 107.2.1
 3. SHEET P-903 THE MED GAS ISOMETRIC IS SHOWING THE
 VACUUM EXHAUST PIPING LABELED AS MV PRESSURE RELIEF
 VENT IT SHALL BE LABELED TO MATCH THE REST OF THE PLAN
 AND THE TEXT IN THE CODE AS VACUUM EXHAUST PIPING. PER
 NFPA 99
 4. PLEASE STATE ON THE PLAN WHAT CATEGORY THE MED GAS
 SYSTEM IS DESIGNED TO BE. PER NFPA 99 CHAPTER 4
 5. THE LIFE SAFETY SHEETS SHOWS A MED GAS STORAGE
 SCHEDULE PLEASE INCLUDE THIS ON THE MED GAS PLAN
 SHEETS. PER WPB AMEND TO FBC 107.2.1
  
  
 PLEASE NOTE THAT SUBMITTAL OF ADDITIONAL AND/OR REVISED
 MATERIALS MAY RESULT IN NEW PLAN REVIEW COMMENTS.
 WHEN RESUBMITTING, IT IS HELPFUL TO PROVIDE A RESPONSE
 LETTER ADDRESSING EACH ITEM ALONG WITH THE CITY
 RE-SUBMITTAL FORM. PLEASE, ADDITIONALLY, INSERT
 CORRECTED PAGES INTO TO SUBMITTAL AND REMOVE OR VOID
 THE PREVIOUSLY REVIEWED SHEETS.
  