2018-10-25 11:05:04 | 1ST REVIEW: |
| 2017 FLORIDA BUILDING CODE 6TH EDITION |
| 2017 FLORIDA BUILDING CODE - PLUMBING, SIXTH EDITION |
| 2017 FLORIDA BUILDING CODE - ACCESSIBILITY, SIXTH |
| EDITION |
| 2017 FLORIDA BUILDING CODE - FUEL GAS, SIXTH EDITION |
| 2017 FLORIDA BUILDING CODE - RESIDENTIAL, SIXTH EDITION |
| 2017 FLORIDA BUILDING CODE - EXISTING BUILDING, SIXTH |
| EDITION |
| 2017 FLORIDA BUILDING CODE - ENERGY CONSERVATION, SIXTH |
| EDITION |
| |
| MED GAS COMMENTS: |
| 1. PROVIDE A SCHEDULE FOR ALL MED GAS EQUIPMENT. PER |
| WPB AMEND TO FBC 107.2.1 |
| 2. PROVIDE MANUFACTURE SPECIFICATIONS AND INSTALLATION |
| GUIDES FOR ALL MED GAS EQUIPMENT INCLUDING THE TANKS |
| AND VALVES. PER WPB AMEND TO FBC 107.2.1 |
| 3. SHT. P-801 DETAIL #4 SHOWS VACUUM EXHAUST |
| DISCHARGING THROUGH THE EXTERIOR WALL IT SHALL |
| DISCHARGE THROUGH ROOF. PER FBC PL 713.6 AND NFPA 99/ |
| 5.1.3.7.7 |
| 4. SHT. P-903 ON THE ISOMETRIC WHEN A PIPE IS CROSSING |
| ANOTHER PIPE IT SHOULD HAVE A GAP AS THEY LOOK NOW A |
| NUMBER OF PIPES APPEAR TO BE CONNECTED TOGETHER IT IS |
| DIFFICULT TO DETERMINE WHICH SYSTEMS IS WHICH PLEASE |
| CLARIFY. PER WPB AMEND TO FBC 107.2.1 |
| 5. PLEASE PROVIDE DETAIL OF THE STRAPPING OF THE MED |
| GAS STORAGE TANKS. PER WPB AMEND TO FBC 107.2.1 |
| 6. A SEPARATE PERMIT IS REQUIRED FOR MED GAS, ALONG |
| WITH THE APPLICATION PROOF OF MED GAS CERTIFICATION FOR |
| THE CONTRACTOR AND INSTALLER IS REQUIRED. PER FAC |
| 61G4-15.031(1)(4)(A)(B) |
| 7. PROVIDE A RISER DIAGRAM FOR THE VACUUM EXHAUST AND |
| THE PRESSURE RELIEF LINES FOR THE MED GAS SYSTEM. PER |
| WPB AMEND TO FBC 107.2.1 AND NFPA 99/ 5.1.3.7.7, |
| 5.1.3.5.6 |
| 8. CLEARLY SHOW THE TERMINATION POINTE OF THE VACUUM |
| EXHAUST AND RELIEF PIPING ON THE ROOF PLAN. PER WPB |
| AMEND TO FBC 107.2.1 AND NFPA 99/ 5.1.3.7.7, 5.1.3.5.6 |
| |
| |
| PLEASE NOTE THAT SUBMITTAL OF ADDITIONAL AND/OR REVISED |
| MATERIALS MAY RESULT IN NEW PLAN REVIEW COMMENTS. |
| WHEN RESUBMITTING, IT IS HELPFUL TO PROVIDE A RESPONSE |
| LETTER ADDRESSING EACH ITEM ALONG WITH THE CITY |
| RE-SUBMITTAL FORM. PLEASE, ADDITIONALLY, INSERT |
| CORRECTED PAGES INTO TO SUBMITTAL AND REMOVE OR VOID |
| THE PREVIOUSLY REVIEWED SHEETS. |
| |