Plan Review Notes For Permit 01101533 |
Permit Number |
01101533 |
|
Review Stop |
E |
Sequence Number |
2 |
|
Notes |
Date |
Text |
2001-12-06 00:00:00 | ****************** UNSAT ************ | | | | 1)NOTE: PLEASE NOTE WIRE SIZE FEEDING | | SIZE FEEDING X-RAY MACHINE SHOWN AT #4 | | AND IS RATED FOR 85AMPS. NEXT SIZE OCP | | WOULD BE 90AMPS. PLEASE NOTE THAT | | PANEL SCHEDULE SHOWS 100A OCP. | | | | 2)NOTE: PLEASE NOTE THAT CONDUCTOR SIZES | | FOR 125A SERVICE ARE TOO SMALL PER | | 310-16. SHOWN AS #3. | | | | 3)NOTE: PLEASE NOTE THAT NOW PLANS ARE | | NOW SHOWING A NEW 125A MAIN. PLEASE | | SUBMIT AIC FOR NEW MAIN. PER 110-9 | | | | 4)NOTE: PLEASE NOTE THAT ANY HANDDRAWN | | CHANGES ON PLANS, MUST BE DONE BY THE | | ENGINEER OF RECORD. PLEASE ALSO NOTE | | THAT PLANS NEED TO BE RE-PRINTED. PENCIL | | CHANGES ARE NOT PERMITTED EITHER. | | 417.025/ | | | | | | PLEASE SUBMIT THE ABOVE INFORMATION FOR | | REVIEW. IF THERE ARE ANY QUESTIONS, | | PLEASE CALL. | | | | DEWEY PALMER | | ELECTRICAL PLAN REVIEW | | 561-659-8096 EXT 8372 |
|