2001-03-21 00:00:00************DENIED**********************
 1. ALL MED GASSHALL COMPLY WITH 1999
 NFPA 99C
 2. SPECIFY PIPE MATERIALS ON PLAN.
 VACUUM PIPING TO COMPLY WITH 4-3.2.2.2
 OXYGEN PIPING BE PER 4-3.1.2.7
 3. SPECIFY PIPE SIZES ON PLAN FOR O2 AND
 VACUUM SYSTEMS FOR REVIEW.
 4. NOTE ON P-4 REFERS TO P-1 FOR VACUUM
 PUMP PIPING AND VALVE CONFIGURATION.
 P-1IS NOT INCLUDED FOR REVIEW.
 5. THE MAIN VALVE (O2) IS SHOWN LOCATED
 AFTER THE FIRST BRANCH WHICH IS SHOWN
 CONNECTED BACK TO THE REST OF THE
 SYSTEM IN THE CORRIDOR. MAIN VALVE
 NEEDS TO BE PRIOR TO ANY BRANCH.
 6. PIPING IS TO BE PROMINENTLY LABELED
 AS TO USE PRIOR TO INSTALLATION PER
 4-3.2.2.2
 7. PRESSURE TEST PER 4-3.4.1.1 SHALL BE
 BEFORE WALLS ARE CLOSED.
 8. THE EMERGENCY OXYGEN SUPPLY
 CONNECTION SHALL BE LOCATED ON THE
 EXTERIOR OF THE BUILDING WHERE IT IS
 ACCESSIBLE BY EMERGENCY SUPPLY VEHICHLE
 AT ALL TIMES.
 9. NO AREA ALARMS ARE SHOWN FOR THE
 PROCEDURE ROOMS. PROVIDE INFORMATION ON
 AREA ALARMS PER 4-3.1.2.2 (C)
 10. ALARM PANELS SHALL BE POWERED BY THE
 LIFE SAFETY BRANCH OF THE EMERGENCY
 SYSTEM PER 4-3.1.2.2
  
 PROVIDE PLANS INCLUDING INFORMATION
 ABOVE FOR REVIEW.
 CONTACT PAUL SCHMITZ CHIEF PLUMBING
 INSPECTOR @ (561) 659-8096 EXT. 8233
 FOR REQUIRED INSPECTIONS AND TESTS
 REQUIRED BY
 THIS OFFICE FOR THIS PERMIT TYPE.