Plan Review Notes
Plan Review Notes For Permit 99090628
Permit Number 99090628
Review Stop E
Sequence Number 1
Notes
Date Text
2001-01-08 00:00:001) SUBMIT PANEL SCHEDULES FOR A,B,C
  
 2) SUBMIT RISER DIAGRAM
  
 3) NO BRANCH CIRCUIT NUMBERS SHOWN ON
 PLAN!
  
 4) PLANS SHOW NO REFERENCE TO 517 FOR
 RECEPTACLE PLACMENT!!!!!
  
 5) 517-18, 517-19, 517-33 FOR PATIENT
 CARE AREAS!!!! IF THESE DO NOT APPLY
 WHY?????????????????????????????????