2001-01-08 00:00:00**************PROVISO*****************
 1) DENTAL VACUUM AND COMPRESSED AIR TO
 BE INSTALLED AND TESTED UNDER 1996 NFPA
 99C
 2) MECHANICAL PERMIT REQUIRED FOR INSTAL
 LATION OF ABOVE SYSTEM. PROVIDE DETAILED
 PLANS SHOWING COMPLIANCE WITH NFPA 99C
 I.E 4-5.2.2.3 REQUIRES ACCESSIBLE CLEAN-
 OUTS FOR VACUUM PIPING AND 4-5.1.1.3(E,F
 G AND H) INFORMATION NOT SHOWN ON PLAN
 3) 1" PVC FRESH AIR LINE NOT SHOWN ON
 HVAC PLAN AS STATED ON P-2 DETAIL 1-P-2
 4) HORN AND STROBE REQUIRED FOR NOTIFI-
 CATION OF FAN SHUT DOWN BY DUCT SMOKE
 DETECTOR IF NOT TIED TO BLDG FIRE ALARM
 PER NFPA 90A 1996 4-4.4.2 AND 4-4.4.3