Plan Review Notes
Plan Review Notes For Permit 20121298
Permit Number 20121298
Review Stop P
Sequence Number 1
Notes
Date Text
2021-01-22 10:36:501) SUBMIT ADDITIONAL INFORMATIONL FOR THE STERILIZER
 SUCH AS MANUFACTURER'S SPECIFICATIONS AND INSTALLATION
 MANUAL. WPB AMEND. TO FBC SEC. 107.2.1.
 2) A BACKFLOW PREVENTION DEVICE WILL BE REQUIRED ON THE
 COLD WATER SUPPLY TO THE STERILIZER IF THE SUPPLY IS
 POTABLE WATER. CLARIFY THE ACRONYMN "NPCW."
 FBC-PLUMBING SEC.608.3.1.
 3) INDICATE THE TYPE OF PIPE FOR THE DRAIN ON THE
 STERILIZER. WHERE THE WASTE WATER TEMPERATURE WILL BE
 GREATER THAN 140 DEGREES, THE PIPING MATERIAL SHALL BE
 RATED FOR THE HIGHEST TEMPERATURE OF THE WASTE WATER.
 FBC-PLUMBING SEC. 702.5.
 4) WHAT IS THE TEMPERATURE OF THE HOT WATER? THE
 THICKNESS OF THE REQUIRED INSULATION SHALL COMPLY WITH
 FBC-ENERGY TABLE C403.2.10.REFER TO SHT. P1.1 -
 PLUMBING SPECIFICATIONS NOTE #14 (C) -
 5) SUBMIT THE INSTALLATION MANUAL FOR THE STERILIZER.
 WPB AMEND. TO FBC SEC. 107.2.1.
 6) SHT. P1.1 - PLUMBING NOTES - NOTE #6 REFERS TO 1/2 "
 NPHW RINSE WATER LINE BUT THIS LINE IS NOT SHOWN ON THE
 STERILIZER PIPING DIAGRAM. CLARIFY.
 7) PLUMBING SHEET NOT DIGITALLY SIGNED AND SEALED AND
 REQUIRES 3RD PARTY CERTIFICATION PER FL. ADDMIN. CODE
 RULE 61G15-23.004.
  
 TIM LARGE
 CHIEF PLUMBING INSPECTOR
 PLUMBING PLAN REVIEW
 561-805-6692
 [email protected]
  
  