2019-05-01 13:39:54PLUMBING 1ST REVISION REVIEW**DENIED**
  
 1) PLAN'S THAT CONTAIN MED GAS PIPING IS SO CONGRESSED
 THAT IT IS IMPOSSIBLE TO PERFORM AN ACTUATE PLAN
 REVIEW. PROVIDE FLAT DRAWINGS OF ALL MED GAS PIPING IN
 SMALLER SECTION ON FULL SHEETS AND PROVIDE A COMPLETE
 ISOMETRIC DRAWING TO INCLUDE DETAIL PIPING OF ALL NEW
 MED GAS EQUIPMENT. FBC 107.3.5.
  
 2) PROVIDE COMPLETED INSTALLATION MANUALS FOR ALL NEW
 MED GAS EQUIPMENT. FBC 107.3.5
  
 3) CERTIFICATION FOR THE PLUMBING CONTRACTOR AND ANY
 AND ALL EMPLOYEES SHOWING MED GAS CERTIFIED IS REQUIRED
 TO BE SUBMITTED PRIOR TO THE MED GAS PERMIT BEING
 ISSUED. FBC 105.1
  
  
 PLEASE NOTE THAT SUBMITTAL OF ADDITIONAL AND/OR REVISED
 MATERIALS MAY RESULT IN NEW PLAN REVIEW COMMENTS.
 WHEN RESUBMITTING, IT IS HELPFUL TO PROVIDE A RESPONSE
 LETTER ADDRESSING EACH ITEM ALONG WITH THE CITY
 RE-SUBMITTAL FORM. PLEASE, ADDITIONALLY, INSERT
 CORRECTED PAGES INTO TO SUBMITTAL AND REMOVE OR VOID
 THE PREVIOUSLY REVIEWED SHEETS.
  
  
  