Plan Review Notes
Plan Review Notes For Permit 19021006
Permit Number 19021006
Review Stop MEDGAS
Sequence Number 1
Notes
Date Text
2019-03-28 07:59:261ST REVIEW: 19021006
 2017 FLORIDA BUILDING CODE 6TH EDITION
 2017 FLORIDA BUILDING CODE - PLUMBING, SIXTH EDITION
 2017 FLORIDA BUILDING CODE - ACCESSIBILITY, SIXTH
 EDITION
 2017 FLORIDA BUILDING CODE - FUEL GAS, SIXTH EDITION
 2017 FLORIDA BUILDING CODE - RESIDENTIAL, SIXTH EDITION
 2017 FLORIDA BUILDING CODE - EXISTING BUILDING, SIXTH
 EDITION
 2017 FLORIDA BUILDING CODE - ENERGY CONSERVATION, SIXTH
 EDITION
  
 MED GAS COMMENTS:
  
 1. PROVIDE ROOF DRAWING SHOWING DISTANCE FROM VACUUM
 EXHAUST TO AIR INTAKES. PER NFPA 99
 2. PLEASE STATE ON PLAN WHAT CATEGORY THE MED GAS
 SYSTEM IN BEING DESIGNED FOR. PER NFPA 99
 3. PROVIDE NOTE ON THE PLAN THE CONTRACTOR AND
 INSTALLER SHALL BE MED GAS CERTIFIED AND SHALL PROVIDE
 CERTIFICATION FOR PERMIT ISSUANCE AND INSPECTIONS. PER
 FAC 61G4-15.031(1)(4A)(4B)
 4. PROVIDE ISOMETRIC DRAWING FOR MED GAS INCLUDING
 VACUUM EXHAUST. WPB AMEND TO FBC 107.3.5.1.3(13)
 5. PROVIDE PIPING DETAIL OF CONTROL VALVES AND TANK
 MANIFOLD ALSO DETAIL OF HOW TANK ARE SECURED. PER WPB
 AMEND TO FBC 107.2.1 AND NFPA 99
 6. NOTE ON THE PLAN THE NUMBER AND SIZE OF TANKS. PER
 WPB AMEND TO FBC 107.2.1 AND NFPA 99
 7. PROVIDE MANUFACTURE SPECIFICATIONS AND INSTALLATION
 GUIDES FOR ALL MED GAS EQUIPMENT. PER WPB AMEND TO FBC
 107.2.1 AND NFPA 99
  
 PLEASE NOTE THAT SUBMITTAL OF ADDITIONAL AND/OR REVISED
 MATERIALS MAY RESULT IN NEW PLAN REVIEW COMMENTS.
 WHEN RESUBMITTING, IT IS HELPFUL TO PROVIDE A RESPONSE
 LETTER ADDRESSING EACH ITEM ALONG WITH THE CITY
 RE-SUBMITTAL FORM. PLEASE, ADDITIONALLY, INSERT
 CORRECTED PAGES INTO TO SUBMITTAL AND REMOVE OR VOID
 THE PREVIOUSLY REVIEWED SHEETS.
  
 GEORGE JOHNSON
 PLUMBING PLANS EXAMINER
 CITY OF WEST PALM BEACH
  561-805-6711
 [email protected]