Plan Review Notes
Plan Review Notes For Permit 16050233
Permit Number 16050233
Review Stop E
Sequence Number 2
Notes
Date Text
2016-06-15 08:49:312ND REVIEW 2014 5TH FBC ED. NEC 2011
 MARIO VALDES
 ELECTRICAL PLAN EXAMINER
 561-805-6713
 [email protected]
  
  
 - PLEASE PROVIDE X-RAY EQUIPMENT MANUFACTURE
 SPECIFICATION CUT SHEETS ( 2ND REQUEST )
  
 - LOAD CALCS AND PANEL SCHEDULE REFLECT SINGLE PHASE
 VOLTAGE AND RISER DIAGRAM SHOWS THREE PHASE ? PLEASE
 CLARIFY AND MATCH INFORMATION. ( 2ND REQUEST )