2011-01-13 10:01:20FBC PLUMBING 2007
 NFPA 99C 2002
 REVISION REVIEW #1
 DENIED:
  
 1. PLEASE SUBMIT VACUUM PUMP MANUFACTURERS
 SPECIFICATIONS SHOWING THAT 3/4" MAIN AND 1/2" RISERS
 ARE SUGGESTED OR REQUIRED. RAMVAC VACUUM PUMPS REQUIRE
 2" MAINS BELOW FLOOR SLOPED AT 1/4" PER FOOT. VACSTAR'S
 SMALLEST PUMP VS 20 REQUIRES A MINIMUM OF 1" MAIN -
 ETC.
  
 2. SANITARY DRAWINGS APPEAR TO BE ELIMINATING THE
 DRAINS FOR THE CHAIRS IN TREATMENT ROOMS 1 & 3. ARE
 THESE CHAIRS EXISTING?
  
 3. NFPA 99C 2002 5.3.3.6.3 VACUUM DRAINS ARE TO BE
 DIRECTLY CONNECTED TO THE SANITARY, TRAPPED, AND
 VENTED.
  