2001-03-20 00:00:00PROVISO:
 THIS INSTALLATION SHALL COMPLY WITH
 NFPA 99C 1999 EDITION.
  
 PIPE SHALL BE PROMINENTLY LABELED PRIOR
 TO INSTALLATION PER 4-3.2.2.2
  
 PRESSURE TEST PER 4-3.4.1.1 SHALL BE
 PRIOR TO WALLS BEING CLOSED.
  
 ALARM PANELS SHALL BE POWERED BY LIFE
 SAFETY BRANCH OF THE EMERGENCY SYSTEM
 PER 4-3.1.2.7
  
 PIPE MATERIAL TO BE PER 4-3.1.2.7 (C)
  
 PIPING SHOWN AS MEDICAL EVACUATION ON
 THIS PLAN IS TO BE INSTALLED PER 4-3.3
 WASTE ANESTHETIC GAS DISPOSAL SYSTEMS
 OR SUBMIT REVISION DEFINING USE OF THIS
 SYSTEM FOR REVIEW.
  
 CALL FOR INSPECTION PRIOR TO CONCEALING
 WORK. CALL FOR REQUIRED INSPECTIONS AND
 ACCEPTANCE TESTS REQUIRED BY THIS
 OFFICE.
 CALL PAUL SCHMITZ @ (561) 659-8096 EXT
 8233 OR JOHN LEECH EXT 8369 IF YOU HAVE
 ANY QUESTIONS ON REQUIRED TESTS OR
 INSPECTIONS.