| 2014-07-09 06:30:29 | BUILDING PLAN REVIEW |
| | PERMIT: 14061244 |
| | ADD: 4440 BEACON CIRCLE |
| | CONT: STRUCTURE TECH |
| | TEL: (305)233-1100 |
| | |
| | 2010 FLORIDA BUILDING CODE W |
| | * 2010 WEST PALM BEACH AMENDMENTS TO THE FLORIDA |
| | BUILDING CODE, CHAPTER 1, ADMINISTRATION, 2010 EDITION |
| | 2012 FBC SUPPLEMENTS ADOPTED APRIL 25/2013. |
| | |
| | 2010 EXISTING BUILDING CODE LEVEL II 701.3 |
| | COMPLIANCE. ALL NEW CONSTRUCTION ELEMENTS, COMPONENTS, |
| | SYSTEMS, AND SPACES SHALL COMPLY WITH THE REQUIREMENTS |
| | OF THE FLORIDA BUILDING CODE, BUILDING. |
| | |
| | 1ST REVIEW |
| | DATE: WED. JULY 09/2014 |
| | ACTION: DENIED |
| | |
| | 1) ARCHITECTURAL & STRUCTURAL SHEETS DESIGNED BY GE |
| | HEALTHCARE ARE NOT SIGNED NOR SEALED. 2010 WEST PALM |
| | BEACH AMENDMENTS TO THE FLORIDA BUILDING CODE, CHAPTER |
| | 1, ADMINISTRATION, 107.1 THE CONSTRUCTION DOCUMENTS |
| | SHALL BE PREPARED BY A REGISTERED DESIGN WHERE REQUIRED |
| | BY CHAPTER 471, FLORIDA STATUTES& 61G15 FLORIDA |
| | ADMINISTRATIVE CODE OR CHAPTER 481, FLORIDA STATUTES& |
| | 61G1 FLORIDA ADMINISTRATIVE CODE. WHERE SPECIAL |
| | CONDITIONS EXIST, THE BUILDING OFFICIAL IS AUTHORIZED |
| | TO REQUIRE ADDITIONAL CONSTRUCTION DOCUMENTS TO BE |
| | PREPARED BY A REGISTERED DESIGN PROFESSIONAL. |
| | |
| | PLANS, SPECIFICATIONS, REPORTS OR OTHER DOCUMENTS |
| | PREPARED BY THE DESIGN PROFESSIONAL AND BEING FILED FOR |
| | PUBLIC RECORD SHALL HAVE THE SIGNATURE AND SEAL OF THE |
| | DESIGN PROFESSIONAL AFFIXED TO THE DOCUMENT. |
| | FL ADMIN CODE 61G15-23.002 ENGINEERS |
| | FL ADMIN CODE 61G16.003 ARCHITECTS |
| | |
| | 2) THE ARCHITECTURAL & STRUCTURAL SHEETS DESIGNED BY GE |
| | HEALTHCARE HAVE NOT BEEN REVIEWED BY THE DESIGNER OF |
| | RECORD AND THEIR SHOP DRAWING STAMP OF APPROVAL |
| | AFFIXED. FL 61G1-23.015 (2) THE ARCHITECT IS |
| | RESPONSIBLE FOR SUPERVISING AND REVIEWING ALL PROJECT |
| | DATA, REPORTS, SHOP DRAWINGS ETC... |
| | |
| | 3) THE GE HEALTHCARE ARCHITECTURAL SHEET NOTE 67 |
| | INDICATES A 9'-0" X 9'-0" SECTION OF EXTERIOR WALL IS |
| | TO BE REMOVED AND REBUILT AFTER THE OLD EQUIPMENT |
| | REMOVED AND THE NEW EQUIPMENT IS INSTALLED.AGAIN THERE |
| | IS NO DESIGN PROFESSIONAL TAKING RESPONSIBILITY FOR THE |
| | DESIGN. |
| | PLANS, SPECIFICATIONS, REPORTS OR OTHER DOCUMENTS |
| | PREPARED BY THE DESIGN PROFESSIONAL AND BEING FILED FOR |
| | PUBLIC RECORD SHALL HAVE THE SIGNATURE AND SEAL OF THE |
| | DESIGN PROFESSIONAL AFFIXED TO THE DOCUMENT. |
| | FL ADMIN CODE 61G15-23.002 ENGINEERS |
| | FL ADMIN CODE 61G16.003 ARCHITECTS |
| | |
| | |
| | 4) SHEET A3 MRI FOUNDATION AND FLOOR DETAIL, THE SIZE |
| | OF THE THICKNESS (HEIGHT) OF THE FOUNDATION IS NOT |
| | PROVIDED NOR IS THE THICKNESS OF THE FLOOR SLAB. THE |
| | PSI OR MIX TYPE OF CONCRETE FOR THE FOUNDATION AND SLAB |
| | ARE NOT PROVIDED |
| | |
| | 5) WHEN RESUBMITTING PLANS PLEASE INDICATE THE REVISION |
| | & REMOVE & REPLACE ANY PAGES AS NECESSARY. A |
| | TRANSMITTAL LETTER LISTING THE ORIGINAL REVIEW COMMENT |
| | NUMBER, WITH A DESCRIPTION OF THE REVISION MADE, |
| | IDENTIFYING THE SHEET OR SPECIFICATION PAGE WHERE THE |
| | CHANGES CAN BE FOUND WILL HELP TO EXPEDITE YOUR PERMIT. |
| | THANK YOU FOR YOUR ANTICIPATED COOPERATION. |
| | |
| | JAMES A. WITMER CBO |
| | SENIOR COMMERCIAL COMBINATION PLANS EXAMINER |
| | TEL: 561-805-6715 |
| | FAX: 561-805-6676 |
| | E-MAIL: [email protected] |
| | |