Plan Review Details - Permit 13030652
Plan Review Stops For Permit 13030652
Review Stop B BUILDING (STRUCTURAL)
Rev No 1 Status P Date 2013-03-29 Cont ID  
Sent By jwitmer Date 2013-03-29 Time 08:07 Rev Time 0.00
Received By jwitmer Date 2013-03-28 Time 16:04 Sent To  
Notes
***NONE***

Review Stop E ELECTRICAL
Rev No 1 Status P Date 2013-03-26 Cont ID  
Sent By rlecky Date 2013-03-26 Time 12:08 Rev Time 0.00
Received By rlecky Date 2013-03-26 Time 11:52 Sent To  
Notes
2013-03-26 12:09:15 
 THE FOLLOWING NOTE WAS PLACED ON THE PLANS.
  
 WIRING IN ALL PATIENT CARE AREAS / EXAM ROOMS REQUIRES
 A SPECIFIC LISTED WIRING METHOD THAT PROVIDES REDUNDANT
 GROUNDING AND INCLUDES AN INSULATED COPPER EQUIPMENT
 GROUND. NEC 517.13
  
  

Review Stop FIRE FIRE DEPARTMENT
Rev No 1 Status P Date 2013-03-25 Cont ID  
Sent By wjolin Date 2013-03-25 Time 13:53 Rev Time 0.00
Received By wjolin Date 2013-03-25 Time 13:53 Sent To  
Notes
***NONE***

Review Stop I INCOMING/PROCESSING
Rev No 1 Status N Date 2013-03-29 Cont ID  
Sent By jwitmer Date 2013-03-29 Time 08:08 Rev Time 0.00
Received By jwitmer Date 2013-03-22 Time 17:52 Sent To  
Notes
***NONE***